Sicuro Bioengineering Limited · Pan-African manufacturer and supplier of in vitro diagnostics

Get a quote

Chemistry analysers

Q-Lyte 60 Plus

Electrolyte analyser using direct ion selective electrodes (ISE), with automatic sampling, washing and calibration.

Product line
Chemistry analysers
Documents
1 file
Reagents listed
7
Set-up
Installation and training available

Fig. 01Q-Lyte 60 Plus · 2 photos

01Overview

Q-Lyte 60 Plus is a fully automated electrolyte analyser using direct ion selective electrodes (ISE). It samples, washes and calibrates automatically, and is a closed system that runs on its original reagents, shows the remaining reagent volume in real time and charts quality control results on a Levey-Jennings (L-J) graph.

Test combinations

  • K⁺, Na⁺
  • K⁺, Na⁺, Cl⁻
  • K⁺, Na⁺, Cl⁻, iCa²⁺
  • K⁺, Na⁺, Cl⁻, Li⁺
  • K⁺, Na⁺, Cl⁻, iCa²⁺, TCa, pH
  • K⁺, Na⁺, Cl⁻, iCa²⁺, pH, TCa, TCO₂, AG, KT

02Specifications

Performance
Throughput60 samples/hour
PrincipleISE (direct ion selective electrode)
Test combinations6 configurations, from K⁺/Na⁺ up to K⁺, Na⁺, Cl⁻, iCa²⁺, pH, TCa, TCO₂, AG, KT
Measuring range
K⁺0.3–20.0 mmol/L
Na⁺20.0–200.0 mmol/L
Cl⁻20.0–200.0 mmol/L
iCa²⁺0.1–6.0 mmol/L
Li⁺0.2–5.0 mmol/L
pH3.0–11.0
Sample
Sample typeSerum and others (see IFU)
Sample volume100–200 µL
System
Display5.6-inch colour touch screen
AutomationAutomatic sampling, washing and calibration
CalibrationAutomatic or on demand
ReagentsClosed reagents, real-time reagent volume monitoring
PrinterInternal thermal printer
Result storageUp to 10,000 results
InterfaceRS-232, LIS supported
Quality control
QC managementL-J (Levey-Jennings) graph
QC reagents3-level QC solution
Physical
Dimensions (L × W × H)507 × 390 × 770 mm
Net weight8 kg
PowerAC 100–240 V, 50/60 Hz, 80 W
Working temperature15–30 °C
Humidity20–80%

The brochure has the manufacturer’s full specification.

03Features

  • Direct ISE measurement of K⁺, Na⁺, Cl⁻, iCa²⁺, Li⁺ and pH, depending on configuration
  • Automatic sampling, washing and calibration
  • Real-time display of reagent volume
  • Simple tubing system, easy maintenance

04Reagents and documents

Runs on Q-Lyte electrolyte reagents (3-level QC, Standards A and B) and maintenance reagents (deproteinizer, cleaning solution, PVC electrode activating solution, reference filling solution), listed below.

Reagents for this analyser

Documents

See all chemistry reagents